App DB | Sorbent Technologies, Inc.
Home
sorbtech.com
Back
Contact an Account Manager
Simultaneous determination of furosemide and amiloride in plasma using HPLC with fluorescence detection
Summary
Type:
HPLC
Application #:
104990
Detection:
fluorescence, excitation 360nm, emission 413nm
Conditions:
eluent: acetonitrile - 0.125M Na dodecyl sulphate - 0.01M perchloric acid pH 2.0 (234.6:35:665, w/w)
flow rate 0.6ml/min
Authors:
Reeuwijk HJEM
Source:
J Chromatogr 575 (1992) 269 - 274
Details
Legend: Fig. 2. Chromatogram of an extract of plasma from a volunteer receiving an oral dose of 40mg furosemide and 5mg amiloride A) before administration, B) 360min after administration, 1. furosemide (46n